guidance and Manufacturer’s declaration-electromagnetic immunity

The Pain Relief Patch BlueTouch is intended for use in the electromagnetic environment specified

below. The customer or the user of the Pain Relief Patch BlueTouch should assure that it is used in the

following environment:
Immunity Test

IEC 60601-1-2

Test Level

Compliance

Level

Electromagnetic Environment - guidance

RF Common

mode/

Conducted

Susceptibility

IEC 61000-4-6

Radiated RF

Electromagnetic

Field

IEC 61000-4-3

3 Vrms

150 kHz

to 80 MHz

3 V/m

80 MHz

to 2,5 GHz

3 V

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any

part of the Pain Relief Patch BlueTouch, including

cables than the recommended separation distance

calculated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

d =1,2 √P

d =1,2 √P 80 MHz to 800 MHz

d =2,4 √P 800 MHz to 2,5 GHz

Where P is the maximum output power rating

of the transmitter in watts (W) according to the

transmitter manufacturer and d is the recommended

separation distance in meters (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,

a

should be less than the compliance level in each

frequency range.

b

Interference may occur in the vicinity of equipment

marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected

by absorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically

with accuracy.

To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey

should be considered. If the measured field strength in the location in which the Pain Relief Patch

BlueTouch is used exceeds the applicable RF compliance level above, the Pain Relief Patch BlueTouch

should be observed to verify normal operation. If abnormal performance is observed, additional

measures may be necessary, such as re-orienting or relocating the Pain Relief Patch BlueTouch

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.