5 technical data, Input for r e sistance thermometer – JUMO 707010 dTRANS T01 (T) HART - 2-Wire Transmitter Operating Manual Benutzerhandbuch

Seite 30

Advertising
background image

12

5 Technical data

Input for r

e

sistance thermometer

De

signation

R

ange limits

Ra

nge

A

c

c

ur

acy

1

Pt 100

-200

to

+850

°C

-100

to

+200

°C

-200

to

+850

°C

±0.2

°C

±0.4

°C

Pt 100 JIS

-200

to

+

649

°C

-100

to

+200

°C

-200

to

+649

°C

±0.2

°C

±0.4

°C

Pt 500

-200

to

+250

°C

-100

to

+200

°C

-200

to

+250

°C

±0.2

°C

±0.4

°C

Pt 1000

-200

to

+250

°C

-100

to

+200

°C

-200

to

+250

°C

±0.2°C

±0.4

°C

Ni 100

-60

to

+

250

°C

-60

to

+

250

°C

±0.2

°C

Ni 500

-60

to

+

150

°C

-60

to

+

150

°C

±0.2

°C

Ni 1000

-60

to

+

150

°C

-60

to

+

150

°C

±0.2

°C

Connection cir

c

uit

2

-, 3-, or 4-wir

e

Shortest span

10°C

S

e

nso

r lead

re

sistance

- for 3- and

4-w

ir

e

connect

ion

- for 2-wir

e

connection

11

Ω

pe

r condu

c

tor

measuring r

e

sistance +

22

Ω

inter

n

al lead r

e

sistance

S

e

nso

r curr

ent

<

0.6

m

A

Sampling rate

> 1 measur

ement per secon

d

In

put fil

ter

1st or

de

r di

gi

tal fi

lter;

filter con

stant adjustable fr

om 0

to

100

sec

Special featur

es

can also

be pr

og

rammed in °F;

range limits ar

e f

reely pr

ogrammable;

input isolated fr

om output

1

T

h

e accur

a

cy r

e

fers to

the maximum range span.

Advertising