Cnd/r •operation – Weidmuller CNDR - Universal galvanically isolated signal converter Benutzerhandbuch

Seite 14

Advertising
background image

1

en

CND/R

•Operation

General
The CND/R measures conductivity using conductance cell and Pt100
for temperature compensation. It displays the conductivity at the
reference temperature and has two alarm channels.

Conductance(Ω

-1

),Conductivity(Ω

-1

cm

-1

)andProbefactors

The conductance measurement is the inverse of the resistance
between the two probes. This depends on properties of the probe like
the distance between them. The probe factor P

f

gives the relationship

between measured conductance M

c

and the actual conductivity C

measurement, so that
C = M

c

x P

f

Normally electrical conductivity is measured at 25 °C. Since the
relationship to temperature is frequently non­linear, the CND/R allows
you to enter four additional points for linearisation. The points are
entered as temperature (T) and the ratio (R) of the conductivity at that
temperature (C

T

) to the conductivity at reference temperature (C

25 °C

).

So that
R = C

T

/

C

25 °C

Viewingthecelltemperature
Press the

ì

key to display the cell temperature. If the display shows

Ovr , then the RTD is not connected correctly.

Over-rangedisplay
When the input signal is outside the programmed input range the