Inputcalibrationprocedure – Weidmuller CNDR - Universal galvanically isolated signal converter Benutzerhandbuch

Seite 19

Advertising
background image

1

Equipmentrequirements
• Suitable precision 50 Ω, 200 Ω, 1 kΩ, 5 kΩ resistors (±0.05 % or

better accuracy)

• An Rtd simulator with ±0.05 % or better accuracy
• An accurate digital multimeter (accurate to 0.05 mV and ±0.1 μA)

Terminalconnectionsforinputcalibration

CalibrationStage

Signaltype

Terminal

Calibrating resistance

2
3

Pt100 RTD simulator

A

4

B

5

Bsense

6

Inputcalibrationprocedure

Whenthe

displayshows

Action/Description

ENT

ï

Put the instrument in setup mode

1.10

While the software version number is flashing,
remove the security link and press

ENT

rlay

Display indicates that unit is /R type, press

ENT

CIPn

Press Å or ì to Calibrate the inputs or ENT to skip

CIPy

Calibrate inputs?

Start

No

Connect the RTD Simulator to the inputs (for connections see over page).

OPEN

Leave inputs 2 & 3 open circuit

Calibrate

Skip

20.00

Connect 50 Ω resistor

Calibrate

Skip

5.00

Connect 200 Ω resistor

Calibrate

Skip

1.00

Connect 1 kΩ resistor

Calibrate

Skip

0.20

Connect 5 kΩ resistor

Calibrate

Skip

rCLy

Calibrate RTD Sensor?

Start

No

0C

Set the RTD Simulator to 0 °C

Calibrate

Skip

200C

Set the RTD Simulator to 200 °C

Calibrate

Skip

COPn

Press Å or ì to Calibrate the outputs or ENT to skip to the next setup item