Setup, Calibration – Weidmuller CNDR - Universal galvanically isolated signal converter Benutzerhandbuch

Seite 18

Advertising
background image

18

•Setup

Changingthesetup
For setup mode, connect the security link and press the

Å

and

ì

keys simultaneously. The software version will be displayed (flashing). If
you wish to continue, press within 10 s.
During setup the

Å

and

ì

keys change values and toggle options.

The ENT key accepts changes and moves to the next item in the setup
sequence (see table).
Note: If you remove the security link at any time during the setup
sequence, any changes you have made will be discarded and the
instrument will reset.

•Calibration

General
The CND/R is factory calibrated. Once calibrated it will operate with
any input without recalibration.
Allow the instrument 15 minutes of powered operation (to reach a
stable temperature) before calibration and allow a few seconds for
signals to stabilise before accepting their value.