Baumer Electric AG

35/40

www.baumerelectric.com

Frauenfeld, Switzerland

DACU/I BA V2.1M

6.2.3

Operate, Reset

Turns the DACU on Reset resp. Operate.

Command data structure:

Start

command

Parameter

Checksum

02h

62h

31h

35h

STX

b

1

5 (95h)

Parameter

0:

Reset

1:

Operate

6.2.4

Fix Range

Measuring range setting

Command data structure
Example: 50’000pC

Start

command

P. 1

P. 2

P. 3

Checksum

02h

63h

30h

30h

35h

41h

STX

c

0

0

5

A(FAh)

Parameter 1

0:

CH1

Parameter 2 & 3

01: Range 1

1'000'000 pC

02: Range 2

500'000 pC

03: Range 3

200'000 pC

04: Range 4

100'000 pC

05: Range 5

50'000 pC

06: Range 6

20'000 pC

07: Range 7

10'000 pC

08: Range 8

5'000 pC

09: Range 9

2'000 pC

10: Range 10

1'000 pC

11: Range 11

500 pC

12: Range 12

200 pC

13: Range 13

100 pC

6.2.5

Variable Range

Setting a variable between 100pC and 1'000’000pC

Command data structure:
Example: (0)23’500pC

Start command P. 1

P. 2

P. 3

P. 4

P. 5

P. 6

P. 7

Checksum

02h

64h

30h

30h

32h

33h

35h

30h

30h

30h

STX

d

0

0

2

3

5

0

0

0 (1C0h)

Parameter 1

0:

CH1

Parameter 2 - 7

Numerical value ASCII-Format

Baumer Electric AG

36/40

www.baumerelectric.com

Frauenfeld, Switzerland

DACU/I BA V2.1M

6.2.6

Setting alarm outputs

Alarm output threshold setting.
Command data structure:
Example: Alarm 2 auf (0)8’500mV

Start

command

P. 1

P. 2

P. 3

P. 4

P. 5

P. 6

Checksum

02h

65h

31h

30h

38h

35h

30h

30h

35h

STX

e

1

0

8

5

0

0

5 (195h)

Parameter 1

0:

Alarm 1

1:

Alarm 2

Parameter 2 - 6

Numerical value in ASCII-Format

6.2.7

80% Test

Test function to check the measuring chain.
Output value 8V

Command data structure:
Example: Test on

Start

command

P. 1

Checksum

02h

66h

31h

39h

STX

f

1

9 (99h)

Parameter 1

0:

Test off

1:

Test on

6.2.8

Signal

Read out of the actual measuring value.

Command data structure:

Start

Command

P. 1

Checksum

02h

67h

31h

41h

STX

g

1

A (9Ah)

Parameter 1

0:

Read out of measuring value CH1

1:

Continuous read out of measuring values CH1


Comment:

The continuous read out oft the measuring values can be stopped by changing the parameter 1 from
1 to 0! (Parameter 1 = 0)

6.2.9

Baud rate

The DACU send the answer (ACK;06h) with the previous baud rate back to the master and changes
afterwards to the new baud rate.

Command data structure:
Example: 38’400Baud

Start

command

P. 1

Checksum

02h

68h

37h

31h

STX

h

7

1 (A1h)