6.2.3
Operate, Reset
Turns the DACU on Reset resp. Operate.
Command data structure:
Start
command
Parameter
Checksum
02h
62h
31h
35h
STX
b
1
5 (95h)
Parameter
0:
Reset
1:
Operate
6.2.4
Fix Range
Measuring range setting
Command data structure
Example: 50’000pC
Start
command
P. 1
P. 2
P. 3
Checksum
02h
63h
30h
30h
35h
41h
STX
c
0
0
5
A(FAh)
Parameter 1
0:
CH1
Parameter 2 & 3
01: Range 1
1'000'000 pC
02: Range 2
500'000 pC
03: Range 3
200'000 pC
04: Range 4
100'000 pC
05: Range 5
50'000 pC
06: Range 6
20'000 pC
07: Range 7
10'000 pC
08: Range 8
5'000 pC
09: Range 9
2'000 pC
10: Range 10
1'000 pC
11: Range 11
500 pC
12: Range 12
200 pC
13: Range 13
100 pC
6.2.5
Variable Range
Setting a variable between 100pC and 1'000’000pC
Command data structure:
Example: (0)23’500pC
Start command P. 1
P. 2
P. 3
P. 4
P. 5
P. 6
P. 7
Checksum
02h
64h
30h
30h
32h
33h
35h
30h
30h
30h
STX
d
0
0
2
3
5
0
0
0 (1C0h)
Parameter 1
0:
CH1
Parameter 2 - 7
Numerical value ASCII-Format
Baumer Electric AG
36/40
www.baumerelectric.com
Frauenfeld, Switzerland
DACU/I BA V2.1M
6.2.6
Setting alarm outputs
Alarm output threshold setting.
Command data structure:
Example: Alarm 2 auf (0)8’500mV
Start
command
P. 1
P. 2
P. 3
P. 4
P. 5
P. 6
Checksum
02h
65h
31h
30h
38h
35h
30h
30h
35h
STX
e
1
0
8
5
0
0
5 (195h)
Parameter 1
0:
Alarm 1
1:
Alarm 2
Parameter 2 - 6
Numerical value in ASCII-Format
6.2.7
80% Test
Test function to check the measuring chain.
Output value 8V
Command data structure:
Example: Test on
Start
command
P. 1
Checksum
02h
66h
31h
39h
STX
f
1
9 (99h)
Parameter 1
0:
Test off
1:
Test on
6.2.8
Signal
Read out of the actual measuring value.
Command data structure:
Start
Command
P. 1
Checksum
02h
67h
31h
41h
STX
g
1
A (9Ah)
Parameter 1
0:
Read out of measuring value CH1
1:
Continuous read out of measuring values CH1
Comment:
The continuous read out oft the measuring values can be stopped by changing the parameter 1 from
1 to 0! (Parameter 1 = 0)
6.2.9
Baud rate
The DACU send the answer (ACK;06h) with the previous baud rate back to the master and changes
afterwards to the new baud rate.
Command data structure:
Example: 38’400Baud
Start
command
P. 1
Checksum
02h
68h
37h
31h
STX
h
7
1 (A1h)