Configuration 7 – ALTANA Wave-Scan ROBOTIC Benutzerhandbuch
Seite 16

15
English
Configuration
7
Rating
(3...10.5)
Orange Peel based on ACT panels
Tension
(6...24,5)
Scale for Orange Peel
GM-Tension
(6...21)
GM specification
MB-Tension
(6...20)
Mercedes Benz specification
H-Tension
(0...20)
Honda specification
P-Tension
(6...20)
Honda specification
Hada
(1...6)
Note for Orange Peel (outdated)
CR-overall
(6...100)
Ford correlation (outdated)
OFF
Turns the Scale off.
More information on the various scales
are available in the online help section
of the auto-chart software under File /
New/ Instrument Configuration
(...\scales.htm).
Correction
Defects on the sample surface, such
as scratches or craters, can cause
major errors in measurement values.
When Correction is activated, the
affected scan areas are cut out and
the measurement values are calculated
from the sanitized data. Pressing
“operate” turns Correction on and off.
Plausibility Control
An option for comparing the corrected
and uncorrected measurement value.
The greater the difference between the
corrected and uncorrected data, the
more critical is the surface defect. If
the difference is greater than 20 %, the
measurement will be evaluated as a
faulty measurement. An error message
appears and a new measurement is
then needed.
Pressing “operate” turns Plausibility
Control on and off.
Scanlength
Scanlength can be set to 5, 10 and 20
cm.
Note:
Using a short scanlength will result in
strong statistical skewing of
measurement values for large
wavelengths. Thus, when using the We
or Wd scale with a scanlength of
5 cm, at least 3 measurements should
be made per checkzone to ensure
representative values.
Speed
Speed can be set in the area of 50 to
150 mm/sec.