8testing the instrument – ALTANA Wave-Scan ROBOTIC Benutzerhandbuch
Seite 17
Advertising

16
English
The value printed on the reference tile
is the corrected reading. If the value
measured on the reference tile is within
the printed tolerance range, the
requirements are met.
If the mean value is not within the
desired tolerance range, try carefully
cleaning the test tile. If this produces
no improvement, please contact our
Customer Service department.
Due to the underlying measurement
principle, no calibration of the
instrument is required. It is
recommended, however, to check the
functionality of the instrument at
regular intervals (about once every 3
months). The reference tile included
with delivery is provided for this
purpose.
8
Testing the instrument
Mount the test tile at the position
intended for the instrument test.
Perform a measurement.
Advertising