IEEE 488.2 and Specific Command Reference

400

FTB-500

IEEE 488.2 Required Commands

*TST?

Description

The *TST? query causes an internal self-test and

places a response into the Output Queue

indicating whether or not the device completed

the self-test without any detected errors. Upon

successful completion of *TST?, the device

settings is restored to their values prior to the

*TST?.

Syntax

*TST?

Parameter(s)

None

Response Syntax

<Result>

Response(s)

Result:
The response data syntax for <Result> is

defined as a <NR1 NUMERIC RESPONSE DATA>

element.

The <Result> value ranges from -32767 through

+32767.

A <Result> with a value of zero indicates that

the self-test has been completed without errors

detected. A <Result> with a value not equal to

zero indicates that the self-test was not

completed or was completed with errors

detected.

Example(s)

*TST? returns 0

(self-test was completed with success)