SCPI-Based Errors

FTB-500

427

–222

“Data out of range”

Indicates that a legal program data element was

parsed but could not be executed because the

interpreted value was outside the legal range as

defined by the device (see IEEE 488.2, 11.5.1.1.5).

–223

“Too much data”

Indicates that a legal program data element of

block, expression, or string type was received that

contained more data than the device could handle

due to memory or related device-specific

requirements.

–224

“Illegal parameter

value”

Used where exact value, from a list of possible,

was expected.

–225

“Out of memory”

The device has insufficient memory to perform the

requested operation.

–226

“Lists not same length” Attempted to use LIST structure having individual

LIST's of unequal lengths.

–230

“Data corrupt or stale” Possibly invalid data; new reading started but not

completed since last access.

–231

“Data questionable”

Indicates that measurement accuracy is suspect.

–232

“Invalid format”

Indicates that a legal program data element was

parsed but could not be executed because the data
format or structure is inappropriate. For example

when loading memory tables or when sending a

SYSTem:SET parameter from an unknown

instrument.

Error

Number

Description

Probable Cause