A technische daten, Ftb-720, Ichtig – EXFO iOLM Benutzerhandbuch

Seite 175: Intelligenter optischer strecken-mapper, Technical specifications, General specifications

Advertising
background image

intelligenter optischer Strecken-Mapper

169

A Technische Daten

FTB-720

W

ICHTIG

Änderungen an den nachstehenden technischen Daten sind ohne
Vorankündigung möglich. Die in diesem Kapitel enthaltenen
Informationen dienen nur zur Referenz. Die aktuellen technischen
Daten dieses Produkts finden Sie auf der Website von EXFO auf
www.exfo.com
.

TECHNICAL SPECIFICATIONS

Wavelength (nm)

b

850 ± 20, 1300 ± 20, 1310 ± 20, 1550 ± 20, 1625 ± 15 (filtered)

Dynamic range

(dB)

c, d

27, 26, 36, 34, 34

Event dead zone (m)

e

0.8

Attenuation dead zone (m)

e

4, 4.5, 5, 5, 5

Distance range (km)

Multimode: 0.1, 0.3, 0.5, 1.3, 2.5, 5, 10, 20, 40
Singlemode: 1.25, 2.5, 5, 10, 20, 40, 80, 160, 260

Pulse width (ns)

Multimode: 5, 10, 30, 50, 100, 275, 500, 1000
Singlemode: 5, 10, 30, 50, 100, 275, 500, 1000, 2500, 10 000, 20 000

Launch conditions

f

Class CPR 1 or 2

i

Linearity (dB/dB)

b

±0.03

Loss threshold (dB)

0.01

Loss resolution (dB)

0.001

Sampling resolution (m)

Multimode: 0.04 to 2.5
Singlemode: 0.04 to 5

Sampling points

Up to 256 000

Distance uncertainty (m)

g

±(0.75 + 0.0025 % x distance + sampling resolution)

Measurement time

User-defined (60 min. maximum)

Typical real-time refresh (Hz)

3

Stable source output power (dBm)

h

–3 (1300 nm), –7 (1550 nm)

GENERAL SPECIFICATIONS

Size (H x W x D)

130 mm x 36 mm x 252 mm (5

1

/

8

in x 1

7

/

16

in x 9

15

/

16

in)

Weight

0.65 kg (1.4 lb)

Temperature

Operating
Storage

0 °C to 50 °C (32 °F to 122 °F)
–40 °C to 70 °C (–40 °F to 158 °F)

Relative humidity

0% to 95% non-condensing

NOTES

a.

All specifications valid at 23 °C ± 2 °C with an FC/PC connector, unless otherwise specified;
APC connector for FTB-720 singlemode model.

b. Typical.

c. Typical dynamic range with longest pulse and three-minute averaging at SNR = 1.

d.

Multimode dynamic range is specified for 62.5 μm fiber; a 3 dB reduction is seen when testing
50 μm fiber.

e.

Typical dead zone for multimode reflectance below –35 dB and singlemode reflectance below
–45 dB, using a 5 ns pulse.

f. For multimode port, controlled launch conditions allow 50 μm and 62.5 μm multimode fiber testing.

g. Does not include uncertainty due to fiber index.

h.

Typical output power is given at 1300 nm for multimode output and 1550 nm for singlemode output.

i. Under improvement to achieve better conditions.