Operation

Proline Promass 80 PROFIBUS PA

44

5.6

PROFIBUS PA hardware settings

5.6.1

Hardware write protection

A jumper on the I/O board provides the means of switching hardware write protection on or off.
When hardware write protection is switched on, it is not possible to write to the device functions
via PROFIBUS (acyclic data transmission, e.g. via the operating program "FieldCare").

#

Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.

1.

Switch off power supply.

2.

Remove the I/O board.

→ Page 81 ff.

3.

Configure the hardware write protection accordingly with the aid of the jumpers (see Figure).

4.

Installation is the reverse of the removal procedure.

a0001359

Fig. 29:

Switching write protection on and off with the aid of a jumper on the I/O board

1

Jumper for switching write protection on and off

1.1

Write protection switched off (factory setting) = it is possible to write to the device functions via PROFIBUS
(acyclic data transmission, e.g. via the operating program "FieldCare")

1.2

Write protection switched on = it is not possible to write to the device functions via PROFIBUS (acyclic data
transmission, e.g. via the operating program "FieldCare")

2

Jumper without function

LED Overview of LED states:

– Lit continuously

→ Ready for operation

– Not lit

→ Not ready for operation

– Flashing

→ System or process error present → Page 69 ff.

1

1.2

1.1

2