Proline Promass 80 PROFIBUS PA

Commissioning

47

6.3

Quick Setup

In the case of measuring devices without a local display, the individual parameters and functions
must be configured via the operating program, e.g. FieldCare.
If the measuring device is equipped with a local display, all the important device parameters for
standard operation, as well as additional functions, can be configured quickly and easily by means
of the following Quick Setup menus.

6.3.1

Quick Setup "Commissioning"

a0004749-en

Fig. 31:

Quick Setup "Commissioning"

+

+

+

E

E

HOME-POSITION

Esc

E

+

-

XXX.XXX.XX

Yes

Yes

No

No

Mass flow

Volume flow

Density

Selection System Units

Configuration another unit ?

Automatically configuration display ?

Temperature

Quit

Unit

Mass flow

Unit

flow

Volume

Unit

Density

Unit

Temperature

Unit

Totalizer

Unit

Totalizer

Automatically

parameterization

of the display

Language

Pre-setting

Quick Setup

Quick Setup
Commission

n

o

Corr. Vol. flow

Unit

Corr. Vol. flow

Unit

Ref. Density

Fix

Ref. Density

p

Selection pre-settings

Actual Settings

Deliver Settings

y

m

Set Unit to Bus

SET UNITS

OFF

m

!

Note!

• The display returns to the cell SETUP COMMISSIONING (1002) if you the press

X key combination during

parameter interrogation. The stored parameters remain valid.

• The "Commissioning" Quick Setup must be carried out before one of the Quick Setups explained below is run.

m The "DELIVERY SETTINGS" option sets every selected unit to the factory setting.

The "ACTUAL SETTINGS" accepts the units you configured beforehand.

n Only units not yet configured in the current Setup are offered for selection in each cycle. The unit for mass, volume

and corrected volume is derived from the corresponding flow unit.

o The "YES" option remains visible until all the units have been configured. "NO" is the only option displayed when

no further units are available.

p The “automatic parameterization of the display” option contains the following basic settings/factory settings:

YES: line 1= mass flow; line 2 = totalizer 1

NO: The existing (selected) settings remain.