Commissioning

Proline Promass 80 PROFIBUS PA

62

6.7

Adjustment

6.7.1

Zero point adjustment

All measuring devices are calibrated with state-of-the-art technology. The zero point obtained in
this way is printed on the nameplate.
Calibration takes place under reference operating conditions.

→ Page 92 ff.

Consequently, the zero point adjustment is generally not necessary!

Experience shows that the zero point adjustment is advisable only in special cases:
• To achieve highest measuring accuracy also with very small flow rates.
• Under extreme process or operating conditions (e.g. very high process temperatures or very high

viscosity fluids).

Preconditions for a zero point adjustment

Note the following before you perform a zero point adjustment:
• A zero point adjustment can be performed only with fluids that have no gas or solid contents.
• Zero point adjustment is performed with the measuring tubes completely filled and at zero flow

(v = 0 m/s). This can be achieved, for example, with shutoff valves upstream and/or downstream
of the sensor or by using existing valves and gates.
– Normal operation

→ valves 1 and 2 open

– Zero point adjustment with pump pressure

→ Valve 1 open / valve 2 closed

– Zero point adjustment without pump pressure

→ Valve 1 closed / valve 2 open

a0003601

Fig. 35:

Zero point adjustment and shutoff valves

"

Caution!
• If the fluid is very difficult to measure (e.g. containing entrained solids or gas) it may prove

impossible to obtain a stable zero point despite repeated zero point adjustments. In instances of
this nature, please contact your E+H service center.

• You can view the currently valid zero point value using the ZERO POINT function (see the

"Description of Device Functions" manual).