Detail setup, Setting the language, Marking the measuring point (tag) – WIKA UPT-21 Benutzerhandbuch

Seite 44: Setting the tag short, Setting the tag long, Gb 11. detail setup 11. detail setup

Advertising
background image

44

WIKA operating instructions process transmitter, model UPT-2x

14068347.01 06/2014 GB/D

GB

11. Detail setup

11. Detail setup

Requirement: Display and operating unit fitted.

11.1 Setting the language

Available languages: German, English, French, Spanish

1. Open the operating menu with [↵].

Select “Detail setup” and confirm with [↵].

2. Select “Language” and confirm with [↵].

3. Select language and confirm with [↵].

Language is set.

11.2 Marking the measuring point (TAG)

11.2.1 Setting the TAG short

TAG short enables 8 figures with a limited character set (numbers and capital letters).

TAG short can be indicated in the additional display.

1. Open the operating menu with [↵].

Select “Detail setup" and confirm with [↵].

2. Select “Marking” and confirm with [↵].

3. Select “TAG short” and confirm with [↵].

4. Change figure using [▲] [▼] and confirm with [↵].

The cursor moves to the next figure. Repeat for each

figure.

TAG short is set.

11.2.2 Setting the TAG long

TAG long enables 32 figures with alphanumeric characters (all characters in accordance

with HART

®

revision 7). TAG long can be indicated in the additional display.

Setting is made as described in chapter 11.2.1 “TAG short”.

Advertising