Commissioning, operation – WIKA TR10-A Benutzerhandbuch

Seite 21

Advertising
background image

WIKA operating instructions RTD and TC, intrinsically safe designs (Ex i)

3345267.12 01/2014 GB/D

21

GB

Repeated opening/closing is possible; however only if necessary, as it might have a

detrimental effect on the protection class

For cable with a pronounced cold-flow behaviour the screw connection must be fully tightened

6.6 Parallel threads

If the thermometer connecting head, extension neck, thermowell or process connection are

connected with parallel threads (e.g. G ½, M20 x 1.5 ...), these threads must be secured using

seals which prevent liquids from penetrating into the thermometer.

As standard, WIKA uses copper profile seals for the connection between the neck tube and the

thermowell, and flat paper seals for the connection of the connection head and the extension

neck or thermowell.

If the thermometer and the thermowell are already connected, the seals will already be mounted.

The plant operator must check whether the seals are suitable for the operating conditions and

must replace them, if necessary, with suitable seals.

For thermometers without a thermowell, and/or where these are delivered separately, the seals

are not included and must be ordered separately.

Tighten the threads by hand when carrying out the final assembly on the plant. This will

correspond to the delivery status of the premounted components. The final tightening torque

should be applied using a spanner (half rotation).

The seals must be replaced after dismantling!

The seals can be ordered from WIKA, indicating the WIKA order number and/or the

designation (see table).

WIKA

Order No.

Designation

Suitable for

threads

11349981

per DIN 7603 Form C 14 x 18 x 2 -CuFA

G ¼, M14 x 1.5

11349990

per DIN 7603 Form C 18 x 22 x 2 -CuFA

M18 x 1.5, G ⅜

11350008

per DIN 7603 Form C 21 x 26 x 2 -CuFA

G ½, M20 x 1.5

11350016

per DIN 7603 Form C 27 x 32 x 2.5 -CuFA

G ¾, M27 x 2

11367416

per DIN 7603 Form C 20 x 24 x 2 -CuFA

M20 x 1.5

1248278

per DIN 7603 D21.2 x D25.9 x 1.5 -Al

G ½, M20 x 1.5

3153134

per DIN 7603 Form C D14.2 x D17.9 x 2 -StFA

G ¼, M14 x 1.5

3361485

per DIN 7603 Form C D33.3 x D38.9 x 2.5 -StFA

G 1

Legend:

CuFA = Copper, max. 45HB

a

; filled with asbestos-free sealing material

Al

= Aluminium Al99

; F11, 32 to 45 HB

b

StFA = Soft iron, 80 to 95 HB

a

; filled with asbestos-free sealing material

6. Commissioning, operation

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: