Faults, Gb 11. faults – WIKA TR10-A Benutzerhandbuch

Seite 36

Advertising
background image

3345267.12 01/2014 GB/D

36

WIKA operating instructions RTD and TC, intrinsically safe designs (Ex i)

GB

11. Faults

Faults

Causes

Measures

No signal/ line break

Mechanical load too high or

overtemperature

Replace probe or measuring insert

with a suitable design

Erroneous measured

values

Sensor drift caused by

overtemperature

Replace probe or measuring insert

with a suitable design

Sensor drift caused by chemical

attack

Use a design with thermowell

Erroneous measured

values (too low)

Entry of moisture into cable or

measuring insert

Replace probe or measuring insert

with a suitable design

Erroneous measured

values and response

times too long

Wrong mounting geometry, for

example mounting depth too

deep or heat dissipation too high

Temperature-sensitive area of the

sensor must be inside the medium, and

surfaces must be isolated.

Deposits on the sensor or

thermowell

Remove deposits

Erroneous measured

values (of thermocouples)

Parasitic voltages (thermal

voltages, galvanic voltage) or

wrong equalisation line

Use a suitable equalisation line

Indication of the

measured value jumps

Cable break in connecting cable

or loose contact caused by

mechanical overload

Replace probe or measuring insert

with a suitable design, for example

equipped with a strain relief or a thicker

conductor cross-section

Corrosion

Composition of the medium not

as expected or modified or wrong

thermowell material selected

Analyse medium and then select a

more-suitable material or replace

thermowell regularly

Signal interference

Stray currents caused by electric

fields or earth loops

Use screened connecting cables,

increase in the distance to motors and

power lines

Earth circuits

Eliminate potentials, use galvanically

isolated transmitter supply isolators or

transmitters

CAUTION!

If deficiencies cannot be eliminated by means of the measures listed above, shut

down the instrument immediately, and ensure that pressure and/or signal are no

longer present, and secure the instrument from being put back into operation

inadvertently. In this case, contact the manufacturer.

Should a return be necessary, please observe the information in chapter

12.2 “Returns”.

11. Faults

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: