WIKA operating instructions RTD and TC, intrinsically safe designs (Ex i)

GB

6.7 Tapered threads (NPT)

Connections with tapered threads (NPT) are self-sealing and normally must not be sealed. It

should be checked whether it may be necessary to seal them additionally with PTFE tape or

hemp. The threads must be lubricated with a suitable lubricant before fitting.

Tighten the threads by hand when carrying out the final assembly on the plant. This will

correspond to the delivery status of the premounted components. The final tightening and sealing

must be made with a spanner (1.5 to 3 rotations).

7. Information on mounting and operation in hazardous area (Europe)

WARNING!

In hazardous areas the use of a model TR10-A measuring insert without a suitable

connection head (case) is not permissible!

Where required a suitable thermowell is to be used.

7.1 General information on explosion protection

The requirements of the 94/9/EC (ATEX) directive must be followed. Additionally the

specifications of the respective national regulations concerning Ex usage apply.

A) The responsibility for classification of zones lies with the plant operator and not the

manufacturer/supplier of the equipment.

B) The plant operator guarantees, and is solely responsible, that all thermometers in use are

identifiable with respect to all safety-relevant characteristics. Damaged thermometers may not

be used. Repairs may only be completed using original spare parts from the original supplier;

otherwise the requirements of the approval are not fulfilled.

The manufacturer shall not be responsible for constructional modifications after delivery of the

instruments.

C) If a component of electrical equipment, on which the explosion protection depends, is

repaired, then the electrical equipment may only be put back into use, after an authorised

expert has stated that it corresponds to the fundamental characteristics of the requirements

for explosion protection. In addition this expert must provide a certificate for this and provide

the equipment with a test mark.

D) Item C) shall not apply if the component was repaired by the manufacturer in accordance with

the requirements and regulations.