WIKA TR10-A Benutzerhandbuch

Seite 24

Advertising
background image

3345267.12 01/2014 GB/D

24

WIKA operating instructions RTD and TC, intrinsically safe designs (Ex i)

GB

Protective measures for applications that require EPL Ga or Gb:

Operational friction or impact between equipment parts made of light metals or their alloys (e.g.

aluminium, magnesium, titanium or zirconium) with equipment parts made of iron/steel is not

permitted. Operational friction or impact between two light metals is permitted.

7.1.2 Ex marking

For applications without transmitter (digital displays) requiring instruments of equipment Group II

(potentially explosive gas atmospheres), the following temperature class classification and

ambient temperature ranges apply:

Table 1

Marking

Temperature

class

Ambient tempera-

ture range (T

a

)

Max. surface temperature (T

max

)

at the sensor or thermowell tip

II 1G Ex ia IIC T6 Ga

II 1/2G Ex ib IIC T6 Ga/Gb

T6

(-50)

1)

-40 ... +80 °C TM (temperature of the medium) +

self-heating
For this, the special conditions (17)

must be observed.

II 1G Ex ia IIC T5 Ga

II 1/2G Ex ib IIC T5 Ga/Gb

T5

(-50)

1)

-40 ... +95 °C

II 1G Ex iaD IIC T4 Ga

II 1/2G Ex ib IIC T4 Ga/Gb

II 1G Ex ia IIC T3 Ga

II 1/2G Ex ib IIC T3 Ga/Gb

T4, T3

(-50)

1)

-40 ... +100 °C

1) The values in brackets apply to special designs. These sensors are manufactured using special sealing com-

pounds. Moreover, they feature housings made of stainless steel and cable glands for low-temperature ranges

When there is a built-in transmitter and/or a digital display, the special conditions from the type

examination certificate (see item 17) apply.

For applications requiring instruments of equipment Group II (potentially explosive dust

atmospheres), the following surface temperatures and ambient temperature ranges apply:

Table 2

Marking

Power P

i

Ambient

temperature

range (T

a

)

Max. surface temperature

(T

max

) at the sensor or

thermowell tip

II 1D Ex ia IIIC T65 °C Da

II 1/2D Ex ib IIIC T65 °C Da/Db

750 mW

(-50)

1)

-40 ... +40 °C TM (temperature of the medium)

+ self-heating
For this, the special conditions

(17) must be observed.

II 1D Ex ia IIIC T95 °C Da

II 1/2D Ex ib IIIC T95 °C Da/Db

650 mW

(-50)

1)

-40 ... +70 °C

II 1D Ex ia IIIC T125 °C Da

II 1/2D Ex ib IIIC T125 °C Da/Db

550 mW

(-50)

1)

-40 ... +100 °C

1) The values in brackets apply to special designs. These sensors are manufactured using special sealing

compounds. Moreover, they feature housings made of stainless steel and cable glands for low-temperature

ranges.

When there is a built-in transmitter and/or a digital display, the special conditions from the type

examination certificate (see item 17) apply.

7. Information on mounting, operation in hazardous area (Europe)

Advertising