4

Analogueoutputtypeandrange

The output type (current/voltage) is like the input type as it is set by

manufacturing. The output range parameters let the PTX800A show
the output range correctly during review mode and prompt for the
correct values during calibration. They have no affect on the operation
of the instrument. If you change them you must be prepared to
calibrate the outputs.

OutputAction

Direct action results in the normal, proportional output. Reverse action

gives an output that is inverted with respect to the rate, so a rate of
zero gives a full scale output and vice versa.

DisplayIntensity

There are fourteen different levels.

•SetupSequence

Setting

Display

Description

Е

м

PGM

S/W Version

vEr_1.01

S/W Version 1.01

-

-

Next

Inputs

Input type

InP=Cur

InP=voLt

Current (mA) based signal
Voltage based (V) based signal

Toggle

Accept

Input range
low value

InpLO=

4.00

Introduces Input low value
Value, e.g., 4.00 mA

-

Dec

-

Inc

Next

Accept

Input range
High value

InpHI=

20.00

Introduces Input high value
Value, e.g., 20.00 mA

-

Dec

-

Inc

Next

Accept

Input Calibra-
tion?

CALInY

CALInn

Calibrate the inputs
Skip Input calibration

Toggle

Accept

Damping
Factor

dFaCtor=

2

Introduces Input damping factor
Damping Factor , e.g., 2

-

Dec

-

Inc

Next

Accept

Transfer
Function

L nEAr

SQrroot

Linear
Square root

Toggle

Accept