Weidmuller PTX800A - Configurable counter and totalizer Benutzerhandbuch

Seite 45

Advertising
background image

45

ChangetheInputtypeinthesoftware:
1. Power up the instrument and start the setup routine.
. While the software version number is flashing, remove the security

link and press PGM.

. The display will show a number as shown in the table below.

Number

0

1

2

3

4

5

InputType

mA

V

mA

V

mA

V

Outputtype

mA

mA

V

V

None

None

4. Change the number to suit your input and output types.
5. Replace the security link and press PGM.
Changetheoutputrangesettings:
6. Scroll through the setup and change the input type and range

settings - you will be required to calibrate the Inputs.

Calibratetheinput:
. Calibrate the inputs to the new output range and type.

R3

Analogue
Output
Board

R1

R2

D

ispla

y B

oar

d

Main Board

Advertising