Temperature sub range, Clear time tcl for maximum value storage, 8 settings via service interface – LumaSense Technologies IS 210 Benutzerhandbuch

Seite 10: 1 with ht 6000 (accessory), Clear time t, For maximum value storage, Settings via service interface, With ht 6000 (accessory), 8settings via service interface

Advertising
background image

IMPAC pyrometers IS 210 · IGA 210

10

7.4

Temperature sub range

You have the opportunity to choose a sub range (minimum span 51°C)
within the basic measuring range of the pyrometer. This sub range corre-
sponds to the analog output (4 to 20 mA).


7.5

Clear time t

CL

for maximum value storage

The maximum value storage stores the highest measurement value. The
clear time defines the time period until the stored value is deleted and
replaced by a new one.
The following settings are possible:
 Clear time Off: The storage is switched off and only momentary values

are measured.

 If any clear time is set the maximum value storage will be cleared at this time.

 The “auto” mode is used for discontinuous measuring tasks. For example objects are

transported on a conveyer belt and pass the measuring beam of the pyrometer only for a
few seconds. Here the maximum value for each object has to be indicated. In this mode
the maximum value is stored until a new hot object appears in the measuring beam. The
temperature which has to be recognized as “hot“ is defined by the low limit of the adjusted
sub range. The stored maximum value will be deleted when the temperature of the new
hot object exceeds the low limit of the sub range by 1% or at least 2°C.


8

Settings via service interface


8.1

with HT 6000

(accessory)

The HT 6000 is a portable battery driven instru-
ment for the pyrometer parameter settings and
indication of temperature.
The pyrometer has to be connected to the HT
6000 via its service interface cable with the fe-
male SUB-D connector. When the HT 6000 is
switched on it recognizes the pyrometer auto-
matically and the preset parameters can be
displayed. With the “PAR” button the different
parameters can be called. With the arrow buttons

and

the values of the parameters can be

selected and confirmed with the “ENTER button.
If no button is pressed for 30 s the HT 6000
changes to the temperature indication without
accepting the changed value

.

Display

On/Off

Selection of
parameters

Up / Down

Confirmation

SUB-D connector

Settings:

Min

. . .

Max

Settings:

Off

50 ms

. . .

25 s
auto

Advertising