Example for wiring using an external power supply, Connection via usb adapter onto a pc – LumaSense Technologies IS 210 Benutzerhandbuch

Seite 6

Advertising
background image

IMPAC pyrometers IS 210 · IGA 210

6

3.1

Wiring example if using a digital display with integrated power supply:


3.2

Example for wiring using an external power supply:

Note:

Additional analyzing instruments, e.g. controllers, recorders, etc
can be connected in series as shown in drawing above.



3.3

Connection via USB adapter onto a PC (for pyrometer parametrizing)

Controller

°C

Recorder

Digital display

230 V

~

4 ... 20 mA

white

brown

Schirm (schwarz)

24 V DC

Power supply

+ green
- yellow

Power supply
for LED targeting light

White:

+24 V

Pin 1

Brown:

0 V

Pin 2

Black:

Shield

Green:

LED +

Pin 5

Yellow:

LED -

Pin 4

1

2

3

4

5

plug insert

(plug side)

Controller

°C

Recorder

Digital display
with built-in
2-wire supply

230 V

~

white

brown

Shield (black)

4 ... 20 mA

White:

+24 V Pin 1

Brown:

0 V

Pin 2

Black:

Shield

Green:

LED + Pin 5

Yellow:

LED -

Pin 4

+ green

- yellow

Power supply
for LED targeting light

1

2

3

4

5

plug insert
(plug side)

Power supply
as displayed
above

USB adapter

USB port PC

Advertising