Technical data

Temperature range:

250 to 1800°C (482 to 3272°F)

Spectral range:

1.45 to 1.8 µm

Detector:

InGaAs detector

Accuracy:

0.6% of reading

(T

amb

= 23°C, EMI = 100%)

Repeatability:

0.2% of reading

Temperature coefficient: 0.01% of reading in °C per °C deviation of ambient temp. of 23°C
Exposure time t

90

:

20 ms (with dynamical adaptation at low signal levels)

Emissivity:

10 to 100%, adjustable in 1% steps

Temperature display:

3 values / s

large display: 4 digit, height 9 mm
small display: 4 digit, height 4.5 mm

Temperature resolution: on display:

0.1°C: 250 to 999.9°C; 1°C: 1000 to 1800°C
0.1°F: 482 to 999.9°F, 1°F: 1000 to 3272°F

via interface: 0.1°C or 0,1°F

Temp. below 250°C:

on display: 7777

Temp exceeds 1800°C: on display: 8888
Display illumination:

automatically at a certain darkness

Measuring unit:

°C / °F, switchable

Measuring functions:

Normal, MAX, MIN, AVG, DIF

Alarm functions:

Hi-alarm (adjustable)

Sighting:

Laser targeting light (class II, max. laser power
level < 1 mW,

λ = 650 nm)

Data storage:

750 values; storage of: temperature value, date, time, parameter,

emissivity, address, interval, temperature unit in °C or °F

Storage interval:

PRT; OFF; 0.02 s; 0.04 s; 0.1 s; 0.2 s; 0.5 s; 1 s; 2 s; 5 s; 10 s; 20 s; 50s

Serial interface:

USB 2.0

Power supply:

9 V battery (IEC 6LR61) or
9 V-NiMH rechargeable battery (e.g. Varta type 5622)

Battery operation time:

50 hours with 9 V battery (without laser and display illumination)
15 hours with 9 V rechargeable battery (without laser and display
illumination)

Charging connector:

connector for battery charging unit 3 858 490

Ambient temperature:

0 to 60 °C