IMPAC pyrometer IGA 15 plus

7

4

Commissioning


4.1

Battery insertion

For operating the IGA 15 plus a 9 V
battery or 9 V rechargeable battery
is necessary. For insertion of the
battery remove the locking screw
and the battery case cover.
Now the instrument is ready to use.





4.2

Switch on

The instrument’s push button has
two trigger points.


First trigger point

:

The instrument is switched on if the button is pushed to the first trigger point. The measurement
starts after a short self test of approximately 1 second. After releasing the push button the last
measured value will be displayed for 10 seconds (HOLD function) and the sign “HOLD” is shown
on the display. After these 10 seconds the instrument is switched off automatically.
The display indicates “7777” if the measured temperature is below 250°C, it indicates “8888”
above 1800°C.

Second trigger point:
The second trigger point activates the data logger and stores the actual temperature reading. It
can be adjusted if only the actual temperature value is stored or if a continuous storing should be
done as long as the push button is hold on the second trigger point (setting of time interval see
6.10

).

The data logger can store up to 750 temperature values. Should the pyrometer be switched to
single value data logging, the data logging is confirmed by a beep and in the display the used
number of the stored value is indicated for a short time (e.g. for the first stored value STO

#

01 is

shown).

Locking lever:

The push button can be locked in the second trigger point position with the locking

lever by pushing the lever upwards. This activates the continuous measuring mode. Pressing the
trigger once again releases the locking and switches off the continuous measuring mode.


4.3

Sighting

For exact aiming to the object the pyrometer is equipped with a laser target-
ing light. The laser shows a red flashing circle at the position of the measur-
ing spot on the object. The size of the circle varies depending on the distance
and has approximately the size of the measuring spot.

If the instrument is in operation, i.e. push button is in first or second trigger point, the laser
targeting light function can be activated or deactivated by pressing the laser key. If this
function is activated the laser targeting light is automatically switched on as long as the
trigger button is pushed.

Battery case cover

Battery connector

Locking screw for
battery case cover

Locking lever

Push button with 2
trigger points

Battery case