5 electrical installation, 1 pin assignment at the pyrometer, Electrical installation – LumaSense Technologies IN 5/5 Benutzerhandbuch
Seite 19: Pin assignment at the pyrometer, 5electrical installation

INFRATHERM pyrometers IN 5 · IN 5/5
19
5
Electrical installation
The pyrometers are powered by 24 V DC nominal (possible range: 10 ... 30 V). When connecting the device
to the power supply ensure correct polarity. The power consumption (in this case 4 - 20 mA) is also the
measuring signal. The instrument don’t need any time for starting or preheating and is immediately ready for
operation.
To switch off the instrument, interrupt the power supply or unplug the electrical connector.
To meet the electromagnetic requirements (EMV), a shielded connecting cable must be used. The shield of
the connecting cable has to be connected only on the pyrometer’s side. On side of the power supply (switch
board) the shield must be open to avoid ground loops.
IMPAC offers connecting cables, they are not part of standard scope of delivery (see section 13, Reference
numbers).
Example for wiring using a digital display with integrated power supply:
Example for wiring using an external power supply:
Note: Additional analyzing instruments, e.g. controllers, recorders, etc
can be connected in series as shown in drawing above.
5.1
Pin assignment at the pyrometer
Pin
Cable color Connection
Pin 1
(white)
+24 V DC (10 ... 30 V)
Pin 2
(brown)
0 V
Pin 3
(black)
Shield
Instrument’s plug:
3 pin flange connector.
2
3
1
Controller
°C
Recorder
Digital display
with built-in
2-wire supply
230 V
~
white
brown
Shield (black)
4 ... 20 mA
Controller
°C
Recorder
Digital display
230 V
~
4 ... 20 mA
white
brown
Shield (black)
24 V DC
Power supply