WIKA CPB5000HP Benutzerhandbuch

Seite 18

Advertising
background image

Pressure Balance High-Pressure

GB

CPB 5000 HP


WIKA Operating Instructions Pressure Balance High-Pressure Version 2.3

18

3.2.5 Approaching the pressure value (decreasing pressure)

.




Turn the spindle pump anticlockwise to release pressure in the system.




If the pressure is close to the next test level, make the fine adjustment with the spindle wheel.


Caution:
The piston is lowered very quickly just before equilibrium is achieved.



Caution:
Do not remove masses completely from the piston/cylinder system under
pressure.





Once the spindle pump is turned back completely counter-clockwise, only the initial pressure

remains in the system.




Now the high-pressure shut-off valve can be opened.


Open the high-pressure shut-off valve only if the pressure in the system has
been relieved to initial pressure. For this purpose, always first turn the spindle
pump back completely counter-clockwise.





The initial pressure can be further relieved by carefully opening the low-pressure shut-off valve.




By completely opening the low-pressure shut-off valve, the initial pressure will completely escape

back into the tank.




The system is now completely relieved


Even if you wish to relieve the system from a high pressure completely down to
zero immediately, the spindle pump must first be completely turned back
counter-clockwise. Then open the high-pressure shut-off valve, and
subsequently open the low-pressure shut-off valve. The initial pressure then
escapes completely into the tank. The system is now completely relieved.

Advertising