Design and function, Transport, packaging and storage – WIKA GTU-10 Benutzerhandbuch
Seite 9

9
WIKA operating instructions portable SF
6
gas transfer unit, model GTU-10
14051107.01 06/2013 GB/D
GB
3. Specifications / 4. Design and function / 5. Transport, ...
3. Specifications
Gas flow rate:
1.1 m
3
/h (Option: 1.8 m
3
/h)
Maximum pressure:
40 bar abs.
Pressure reducer, outlet pressure: 0 ... 16 bar
Voltage supply:
AC 230 V, 50 Hz (Option: AC 115 V, 60 Hz)
Fuse with 10 A, slow
Operating temperature:
5 ... 40 °C
Weight:
35 kg (with 1.8 m
3
/h)
30 kg (with 1.1 m
3
/h)
Ingress protection:
IP 20
For further specifications see WIKA data sheet SP 63.07 and the order documentation.
4. Design and function
4.1 Short description
With the GTU-10, the user has the option of a pressure-reduced output to fill a gas
chamber with a preset pressure. From this output, the outlet pressure can be set between
0 ... 16 bar.
If, however, the unreduced output is connected, the GTU-10 has the capability to
compress the gas in the vessel to a maximum of 40 bar, and then automatically switch
off. Thus, the GTU-10 gas transfer unit can liquefy SF
6
gas to reduce volume.
4.2 Scope of delivery
Cross-check scope of delivery with delivery note.
5. Transport, packaging and storage
5.1 Transport
Check the portable SF
6
gas transfer unit for any damage that may have been caused by
transport. Obvious damage must be reported immediately.
Only move the unit using the appropriate materials handling equipment e.g. trolley.
Operation of the unit during transport is not permitted.
During any transport, the unit should be secured in accordance with local load regula-
tions. The applicable local safety regulations should be complied with.