Hach-Lange LZV937 - DR 6000 additional software enzymology tests Benutzerhandbuch

Seite 31

Advertising
background image

English 31

6.

To abort the entry without saving, press CANCEL.
To confirm and complete the entry, press DONE.
To call up the next sample, press NEXT.

7.

Close the dilution factor window with CLOSE.

8.

Close the carousel options window with CLOSE.

Note: The configuration can be changed at any time.

Input of initial sample weight

An individual initial sample weight can be included in the results
calculation for each sample.

To enter the initial weight for a sample, proceed as follows:

1.

Under STORED PROGRAMS, select the required test and open
the test with START.

2.

In the toolbar, select the symbol for the carousel insert. This symbol
opens a window with carousel options.

3.

Select INITIAL SAMPLE WEIGHT.

4.

Select the corresponding sample and open with EDIT.

5.

Enter the initial sample weight in g/100 mL.

6.

To abort the entry without saving, press CANCEL.
To confirm and complete the entry, press DONE.
To call up the next sample, press NEXT.

7.

Close the initial sample weight window with CLOSE.

8.

Close the carousel options window with CLOSE.

Note: The configuration can be changed at any time.

The weight-related result is calculated automatically according to the
following formula:

The concentration of the result is illustrated in g/100 g (initial sample
weight).

Changes to the results display

Depending on the test, the result is calculated and output in many
different variants.

If you would like to customize the results information, proceed as
follows:

1.

Select the required test under STORED PROGRAMS.

2.

Select PROGRAM OPTIONS > EDIT.

3.

Highlight the FORMULA row and select EDIT.

4.

Select row CN: G/100 G SUBSTANCE, for example, and select
EDIT.

5.

Highlight the DISP RESULT: YES row and select EDIT.

6.

Select OFF and confirm with OK.

7.

Close the two subsequent windows with OK.

8.

Confirm the change with SAVE.

9.

Close the program with EXIT.

Note: The configuration can be changed at any time.

Cx =

C

mg/L result

× 10

Cy

initial sample weight

Advertising