English 34
Cholesterol
Procedure for cholesterol
1.
Prepare the test according to the process specification.
2.
Select STORED PROGRAMS > SELECT BY NUMBER.
3.
Enter the number 2203 for the program "Cholesterol". Confirm with
OK.
4.
Press START to start the program.
5.
Set sample blank value 1 to carousel position 2, carousel position 1
remains free.
6.
Set sample 1 to carousel position 3.
7.
Set sample blank value 2 to carousel position 4.
8.
Set sample 2 to carousel position 5 and so forth.
9.
Press READ CAROUSEL.
Extinction is measured in all cuvettes.
The result is indicated in g/L cholesterol and g/100 g cholesterol on
the display.
Program number
2203
Parameter
Cholesterol
R-Biopharm AG order number
Art. no. 10 139 050 035
Results information
g cholesterol / L sample solution
g cholesterol / 100 g sample
Carousel loading
Carousel position 1: Free
Carousel position 2: Sample 1 blank value
Carousel position 3: Sample 1
Carousel position 4: Sample 2 blank value
Carousel position 5: Sample 2
Carousel position 6: Sample 3 blank value
Carousel position 7: Sample 3