Hach-Lange LZV937 - DR 6000 additional software enzymology tests Benutzerhandbuch

Seite 36

Advertising
background image

English 36

Acetic acid (acetate) and acetic acid (S)

Procedure for acetic acid

1.

Prepare the test according to the process specification.

2.

Select STORED PROGRAMS > SELECT BY NUMBER.

3.

Enter the number 2205 for the program "Acetic acid". Confirm with
OK.

Enter the number 2206 for the program "Acetic acid (S)" with creep
monitoring. Confirm with OK.

4.

Press START to start the program.

5.

Set the blank value cuvette to carousel position 1.

6.

Set the sample cuvettes to carousel positions 2-7.

7.

Press READ CAROUSEL.

Extinction 1 is measured in all cuvettes. R1 is shown on the display.

8.

Pipette the reagent into the blank value and sample cuvettes
according to the process specification.

9.

Press READ CAROUSEL.

The 3:00 timer starts. When the time elapses, two short acoustic
signals are sounded.

Extinction 2 is measured in all cuvettes. R2 is shown on the display.

10. Pipette the reagent into the blank value and sample cuvettes

according to the process specification.

11. Press READ CAROUSEL.

The 15:00 timer starts. When the time elapses, two short acoustic
signals are sounded.

Extinction 3 is measured in all cuvettes.

Note: For program 2206 acetic acid (S) with creep monitoring,
extinction 3 is measured six times in order to verify the stability of
the value and to preclude a subsequent creep reaction.

The result is indicated in g/L acetic acid and g/100 g acetic acid on
the display.

Program number

2205

2206

Parameter

Acetic acid

Acetic acid (S)

R-Biopharm AG order number

Art. no. 10 148 261 035

Results information

g acetic acid / L sample solution

g acetic acid / 100 g sample

Carousel loading

Carousel position 1: Blank value

Carousel position 2-7: Samples 1 to 6

Advertising