Checking the volume, Volumen kontrollieren – BrandTech Scientific Transferpette -8/-12 Benutzerhandbuch

Seite 15

Advertising
background image

6. Den zuvor gemessenen Istwert

einstellen. Volumeneinstellknopf
ca. 2 mm (ohne zu drehen) bis zum
Anschlag einschieben.

7. Distanzscheibe wieder über die

Achse schieben.

8. Abwerferhaube (A) montieren.
9. Abwerferhaube durch Drehen der

seitlichen Verschlüsse in Position

sichern.

6. Set to the previously measured

actual value. Push back volume
setting knob to the stop (approx.
2 mm), without turning it.

7. Replace the red distance plate

across the axis.

8. Mount ejector cap (A).
9. Secure ejector cap by turning the

lateral closures into position .

Checking the volume

The instrument should be calibrated at least once a
year. A more frequent interval may be dictated by
individual lab rules, applications and conditions.
The gravimetric volume test of the pipette requires
the following steps (page 16):
(Testing procedures are described e.g., in ISO 8655,
Part 6,  “Testing Instruction”, page 37).

Volumen kontrollieren

Das Gerät sollte mindestens einmal jährlich überprüft
werden. Der Zyklus kann aber den individuellen An-
forderungen angepasst werden.
Die gravimetrische Volumenprüfung der Pipette erfolgt
durch folgende Schritte (Seite 16):
(Das Prüfverfahren ist z. B. in ISO 8655, Teil 6
beschrieben,  „Prüfanweisung“, Seite 37).