Trouble shooting – BrandTech Scientific Transferpette -8/-12 Benutzerhandbuch

Seite 32

Advertising
background image

Problem

Possible Cause

Corrective Action

Pipette tip leaks or
volume too low

- Tip does not sit properly

- Press tip on firmly. Use only suitable

high-quality tips, page 8.

- Damaged V-ring (O-ring) on single

nose cone

- Replace V-ring (O-ring) on nose cones.

“Repairs”, pages 26-28.

- Piston damaged

- Replace piston.

“Repairs”, pages 24, 25.

- Seal dirty, damaged

- “Repairs”, pages 22, 23.

- Instrument is dirty

- Clean instrument.

“Cleaning”, page 19.

- Non-aqueous solutions pipetted

- “Adjustment”, page 14.

Poor sample aspiration

- Single nose cone clogged

- Clean nose cone. Remove cones only;

do not disassemble instrument!
“Cleaning” page 19.

Volume too large

- Pipetting key pressed too far down

into the blow-out stroke before
aspirating liquid.

- Operate pipette properly.

“Operating the pipette”, page 10.

Pipetting key moving
stiffly

- Key movement reduced

- Lubricate piston and centrale guide rod

with the appropriate grease,
pages 13, 34.

Trouble Shooting

- 32 -