6 Commissioning, Operation

14 WIKA Operating Instruction, Model CPH6300

14043054

.01

07

/20

12

G

B/

D

Menu

Parameter Value

Description

or

Set Configuration: General configurations

Unit: Select measuring unit

*,**

Arrow to bar,
mbar

Measuring value is displayed in corresponding unit, the
selectable units depend on connected sensor

Arrow to
“user”

Measuring value is displayed in freely adjustable user-unit

bar, mbar

Unit “user”: base unit

0000, 000.0
… .0000

Unit “user”: decimal point setting (only if user-unit selected)

-19999 ...
19999

Unit “user”: multiplication factor

Sea Level: Sea-level correction

*,**

oFF

Sea-level correction off

on

Sea-level correction on (see6.4.3)

-2000 ... 9999 Sea-level in [m]

*,**

Rate: Measuring rate

*

Slo

Slow measuring rate (4 Hz filtered, low power consumption)

FASt

Fast measuring rate, filtered (1000 Hz)

P.dEt

Peak detection: fast measuring rate, unfiltered (1000 Hz)

Averaging Filter

*

1 … 120

Averaging period in seconds

oFF

Averaging function is deactivated

Auto Power-Off: Select power-off delay

1 ... 120

Power-off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes
place.

oFF

Automatic power-off function deactivated (continuous operation)

Background illumination

(factory settings: 5 s)

oFF

Illumination deactivated

5 … 120

Turn off illumination after 5 … 120s

on

Illumination always on