6 Commissioning, Operation

16 WIKA Operating Instruction, Model CPH6300

14043054

.01

07

/20

12

G

B/

D

Menu

Parameter

Value

Description

or

Set Output: Universal output adjustments

oFF

Interface and analog output off

SEr:

Serial interface activated

dAC:

Analog output activated

01,11..91

Base address for serial interface communication
(only at Out = SEr)

CH 1, CH 2 or

DIF

CH

Channel that should be used for analog output
(only at Out = dAC)

***

sensor dep.

e.g.-5.00..5.00 mbar

Zero-point setting for Out = dAC: Measuring value which
should correspond to output 0 V

sensor dep.

e.g.-5.00..5.00 mbar

Slope setting for Out = dAC: Measuring value which should
correspond to output 1 V

Set Corr: Input adjustment (1-channel version)

*

Zero adjustment/offset of sensors

oFF

No zero adjustment of sensors

sensor dep.

e.g.-5.00..5.00 mbar

The offset of sensor will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.

Slope adjustment of sensors

oFF

No slope adjustment of sensor

-2.000 ... 2.000

The slope of sensor will be changed by this factor [%] to
compensate deviations of probe or measuring device

Set Corr: Input adjustment (2-channel version)

*
***

Zero adjustment/offset of sensors 1

oFF

No zero adjustment of sensors 1

sensor dep.

e.g.-5.00..5.00 mbar

The offset of sensor 1 will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.

Slope adjustment of sensors 1

oFF

No slope adjustment of sensor 1

-2.000 ... 2.000

The slope of sensor 1 will be changed by this factor [%] to
compensate deviations of probe or measuring device

Zero adjustment/offset of sensors 2

oFF

No zero adjustment of sensors 2

sensor dep.

e.g.-5.00..5.00 mbar

The offset of sensor 2 will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.

Slope adjustment of sensors 2

oFF

No slope adjustment of sensor 2

-2.000 … 2.000

The slope of sensor 2 will be changed by this factor [%] to
compensate deviations of probe or measuring device