6 Commissioning, Operation

WIKA Operating Instruction, Model CPH6300 23

14043054

.01

07

/20

12

G

B/

D

c) Clear manual recordings:

If data sets have been stored, they can be deleted with the button “STORE”:

Press for 2 seconds: Call menu “Clear”

Select with

or

Clear nothing (cancel menu)

Clear all recordings

Clear the last recording

Confirm selection and quit menu “Clear”


6.5.2 Automatic Recording with Selectable Cycle Time (“Func CYCL“)
If logger function “Func CYCL” is selected (see “Configuration” ) the device will automatically
record measuring values at intervals of the set cycle time.
The logger‘s cycle time can be set from 1 s to 60 min (see “Configuration”).

1-channel version:

Storable data sets: 10000

A data set contains at slow measuring rate (rAtE SLo):

Measuring value at the time of the recording

“CALC off”:

min-peak, max-peak since the last recording

CALC aktiv”

1)

: Calculated value at the time of the recording and dummy value



At data set contains at fast measuring rate (rAtE FASt and rAtE P.dEt):

Arithmetic average since the last recording

“CALC off”:

min-peak, max-peak since the last recording

“CALC aktiv”

1)

: arithmetic average of calculated value since the last recording

and dummy














1)

CALC-Function: see chapter 6.4.5