6 Commissioning, Operation

WIKA Operating Instruction, Model CPH6300 17

14043054

.01

07

/20

12

G

B/

D

Menu

Parameter

Value

Description

or

Set Alarm: Settings for alarm function

On/No.So

Alarm Sensor 1 on, with/without buzzer

OFF

No alarm function for Sensor 1

Sensor1-Min
... AL.1-Hi

Min-alarm-limit sensor 1 (not if AL.1 = oFF)
Sensor1-Min is lower display range limit of sensor 1

AL.1-Lo ...
Sensor1-Max

Max-alarm-limit sensor 1 (not if AL.1 = oFF)
Sensor1-Max is upper display range limit of sensor 1

On/No.So

Alarm Sensor 2 on, with/without buzzer

OFF

No alarm function for Sensor 2

Sensor2-Min
... AL.2-Hi

Min-alarm-limit sensor 2 (not if AL.2 = oFF)
Sensor2-Min is lower display range limit of sensor 2

AL.2-Lo ...
Sensor2-Max

Max-alarm-limit sensor 2 (not if AL.2 = oFF)
Sensor2-Max is upper display range limit of sensor 2

On/No.So

Alarm on, with/without buzzer (1-channel version) or
Alarm sensor difference on with/without buzzer
(2-channel version)

OFF

No alarm function (1-channel version) or
No alarm function for sensor difference (2-channel vers.)

-19999 …
AL.DIF-Hi

Min-alarm-limit difference (not if AL.DIF = oFF)

***

Sensor-Min
... AL.Hi

Min-alarm-limit sensor (not if AL. = oFF)
Sensor-Min is lower display range limit of sensor

AL.DIF-Lo …
19999

Max-alarm-limit difference (not if AL.DIF = oFF)

***

AL.Lo ...
Sensor-Max

Max-alarm-limit sensor (not if AL. = oFF)
Sensor-Max is upper display range limit of sensor

Set Logger: Settings for logger function

*

Selection of logger function

*

CYCL

Cyclic: cyclic logger

Stor

Store: manual recording

oFF

No logger activated

0:01 ... 60:00 Cycle time in [minutes: seconds] (for cyclic logger)

*

on/oFF

Low-power-logger with low power consumption
(only for cyclic logger and slow measuring rate)

*

Set Clock: Settings for real time clock

HH:MM

Clock: set time hours: minutes

YYYY

Year: set year

TT.MM

Date: set date day. month

rEAd Logg: Read calibration data:
seechapter 6.5.1

(*)

If logger memory contains data sets parameters marked with (*) cannot be called. You have to
clear memory to change these parameters! (see chapter 6.5)

(**)

Parameter can only be called if corresponding sensor is plugged to connection 1. The settings
are applied to the sensor at connection 2 (if any connected).

(***)

Only 2-channel version