7.4

Recalibration


DKD certificate - Certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with time
intervals of approx. 12 months. Every factory recalibration includes, additionally, an extensive free-
of-charge check of all system parameters with respect to their compliance with the specification.
The basic settings will be corrected if necessary.


8 Faults

Display

Description

What to do

or

Err.9

No sensor connected

Switch off device and connect sensor

Connected sensor or device
defective

If 2nd sensor available, check if device is ok.
Return defective device/sensor to
manufacturer for repair

Value extremely out of measuring
range

Check: pressure not within sensor range?

No display or

confused

characters,

device does not

react on keypress

Battery empty

Replace battery

Mains operation: wrong voltage or
polarity

Check power supply, replace it when
necessary

System error

Disconnect battery and power supplies, wait
shortly, then reconnect

Device defective

Return to manufacturer for repair

Err.1

Measured value above allowable
range

Check: pressure not within sensor range?
-> measuring value to high!

Sensor defective

Return to manufacturer for repair

Err.2

Measured value below allowable
range

Check: pressure not within sensor range?
-> measuring value to low!

Sensor defective

Return to manufacturer for repair

Err.3

Display range overflow

Check: value above 19999 -> to high to be
displayed

Err.4

Display range underflow

Check: value below -19999 (Tara?) -> to low

Er.11

Value could not be calculated

Choose different unit

Calculation overflow happened

Choose different unit

Err.7

System error

Return to manufacturer for repair

– – – –

Sensor not present / recognised

reconnect sensor,
during logging: stop the logger and restart it

could not calculate value

suitable sensor / unit combination necessary


CAUTION!
If faults cannot be eliminated by means of the measures listed above, the instrument
must be shut down immediately, and it must be ensured that pressure and/or signal
are no longer present, and it must be prevented from being inadvertently put back into
service. In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter "9.2 Return".