8. Problems/solutions

Problem

Possible cause

solution

Heart rate value flashing in

measurements.

Unstable or weak ECG

signals.

Repeat the measurement

as per the instructions in

these Instructions for use.

Measurements show “EE”

(EE = error).

Insufficient contact pressu-

re or you spoke or moved

during the measurement.

Repeat the measurement

as per the instructions in

these Instructions for use.

The device does not

switch on after pressing

the On/Off button for three

seconds.

The battery is empty.

Charge the device using

the USB port on your com-

puter.

The electrode is touching

your skin but the measure-

ment does not start.

Insufficient pressure.

Press the lower electrode

until you hear or feel a click.

The measurements are

no longer displayed in the

device memory.

The memory capacity of 99

measurements has been

reached. As a result, old

data records are replaced

with new ones.

Download data to your

computer on a regular ba-

sis to save them.