Electromagnetic compatibility information – Beurer ME80 Benutzerhandbuch

Seite 49

Advertising
background image

49

Electromagnetic Compatibility Information

Table 1

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic emissions
The ME80 is intended for use in the electromagnetic environment specified below. The customer or the user

of the ME80 should assure that it is used in such an environment.
Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The ME80 uses RF energy only for its internal func-

tion. Therefore, its RF emissions are very low and

are not likely to cause any interference in nearby

electronic equipment.

RF emissions CISPR 11

Class B

The ME80 is suitable for use in all establishments,

including domestic establishments and those

directly connected to the public low-voltage power

supply network that supplies buildings used for

domestic purposes.

Harmonic emissions IEC 61000-3-2

Class A

Voltage fluctuations/flicker emissions

IEC 61000-3-3

Complies