Table 3

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity
The ME80 is intended for use in the electromagnetic environment specified below. The customer or the user

of the ME80 should assure that it is used in such an environment.
IMMUNITY test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete

or ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

2 kV for power

supply lines

1 kV for input/out-

put lines

2 kV for power

supply ports

1 kV for input/out-

put and commu-

nication ports

Mains power quality should be that

of a typical commercial or hospital

environment.

interruptions and voltage

variations on power supply

input lines IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should be that

of a typical commercial or hospi-

tal environment. If the user of the

ME80 requires continued operation

during power mains interruptions,

it is recommended that the ME80

be powered from an uninterruptible

power supply or a battery.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8 3 A/m, 50/60Hz

3 A/m

Power frequency magnetic fields

should be at levels characteristic of

a typical location in a typical com-

mercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.