Table 4

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the

ME EQUIPMENT or ME SYSTEM
The ME80 is intended for use in an electromagnetic environment in which radiated RF disturbances are con-

trolled. The customer or the user of the ME80 can help prevent electromagnetic interference by maintaining a

minimum distance between portable and mobile RF communications equipment (transmitters) and the ME80 as

recommended below, according to the maximum output power of the communications equipment.

Rated maximum output

power of transmitter

(W)

Separation distance according to frequency of transmitter

(m)

150 kHz to 80 MHz

d = 1.2

80 MHz to 800 MHz

d = 1.2

800 MHz to 2.5 GHz

d = 2.3

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in

meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the

maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-

tion and reflection from structures, objects and people.