Specifications
Models WUC-10, WUC-15 and WUC-16
WUC-10 / WUC-15
WUC-16
Pressure ranges
psi
30
60
100
160
250
350
500
1000
1500
bar
2
4
7
11
17
25
36
70
100
Over pressure safety
1)
psi
120
120
210
320
500
750
1100
2100
3000
Burst pressure
1)
psi
1800
1800
2200
2600
4800
6200
7400
8000
10500
Pressure ranges
psi
2000
3000
5000
bar
145
225
360
Over pressure safety
1)
psi
4200
6600
10000
Burst pressure
1)
psi
10500
10500
10500
Other pressure ranges and pressure units (e.g. MPa, kg/cm
2
) on request
Measuring principle
Metal thin-film sensor
Materials
Wetted parts
■
Pressure connection
316L VIM/VAR
■
Pressure sensor
2.4711 / UNS R30003
Case
304 SS
Particle test
≤ 0.1 µm Particle 0.1 ptc / ft³ according to Semi E49.8
Inboard helium leak test
< 1 x 10
-9
mbar l/sec (atm STD cc/sec) according to Semi F1
Surface finish
Electropolished, average Ra ≤ 0.13 µm (RA 5); max. Ra ≤ 0.18 µm (RA 7) exceeds Semi F19
Dead volume
cm
3
WUC-10 < 1.5, WUC-15 < 1, WUC-16 < 1
Permissible Medium
Special gas / Vapour / Liquid
Power supply U+
U+ in VDC
10 ... 30 with output signal 4 ... 20 mA / 0 ... 5 V
14 ... 30 with output signal 0 ... 10 V
Signal output and maximum
ohmic load R
A
R
A
in Ohm
4 … 20 mA, 2-wire
R
A
≤ (U+ – 10 V) / 0.02 A
0 ... 5 V, 3-wire
R
A
> 5 k
0 ... 10 V, 3-wire
R
A
> 10 k
1) 1 psi = 0.069 bar
7. Starting, operation