11384850.07 GB/D 06/2014

WIKA Operating instructions / Betriebsanleitung WUC-1x

19

Failure

Possible cause

Procedure

Output signal unchanged after change in

pressure

Mechanical overload through overpressure

Replace instrument; if failure reoccurs, consult

the manufacturer *)

Wrong supply voltage or current spike

Replace instrument

No output signal

No/incorrect voltage supply or current spike

Adjust the voltage supply to correspond with

the Operating Instructions *)

Cable break

Check connections and cable

9. Maintenance, accessories / 10. Trouble shooting

GB

9. Maintenance, accessories

WIKA transducers require no maintenance.

Have repairs performed by the manufacturer only.

Accessories: For details about the accessories (e. g. connectors), please refer to WIKA‘s price list, WIKA‘s product catalog

on CD or contact our sales department.

10. Trouble shooting

WARNING!

Open pressure connections only after the system is without pressure!

WARNING!

Take precautions with regard to remaining media in removed transducers. Remaining media in the pressure

port may be hazardous or toxic!

Remove the transducer from service and mark it to prevent it from being used again accidentally, if it becomes

damaged or unsafe for operation.

Have repairs performed by the manufacturer only.

Do not insert any pointed or hard objects into the pressure port for cleaning to prevent damage to the diaphragm

of the pressure connection.

Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the right voltage supply and the right

type of wiring (2-wire/3-wire) has been chosen?