6. Maintenance instructions

HYDRA-Dry pressure measuring systems do not require any maintenance.
Cleaning of the HYDRA-Dry may only be accomplished through rinsing. Any
mechanical cleaning inside the process connection may damage the
separating diaphragms.
The HYDRA double diaphragm system prevents contamination of the process
media even in case of mechanical damage of the wetted diaphragm. In this
case the process media will penetrate the lateral leakage boring. The
instrument must be exchanged as soon as possible.
For safety reasons repairs of already mounted instruments are not possible.

7. Specifications

GB

1) Limited by maximum pressure rating of the process connection see section 3.

Specifications

Pressure ranges

bar

0 ... 2.5

0 ... 4

0 ... 6

psi

0 ... 30

0 ... 60

0 ... 100

Overpressure safety of sensor bar

40

1)

Pressure-reference

relative pressure, not suitable for vacuum

Wetted

parts

PFA or TFM (modified PTFE), no transfer-fluid

Power supply UB

VDC

12 <

UB

30 (18 ... 30 V with HYDRA-Indicator)

Signal output

mA

4 ... 20, 2-wire system

Maximum load R

A

R

A

(U

B

-

12 V) / 0.027 A

with R

A

in Ohm and UB in Volt

(with HYDRA-Indicator R

A

(UB - 18 V) / 0.027 A)

Accuracy

% of span

0.5

Stability per year

% of span

0.25

(at reference conditions)

Temperature-coefficients

% of span / 10 K

0.25 (from 5°C to 60°C)

Ingress protection

IP 67