Design and function – WIKA TR12-M Benutzerhandbuch
Seite 14

GB
14064370.01 08/2013 GB/D
WIKA operating instructions models TR12, TC12
14
Connection head, selectable (example)
Thread
Tightening torques in Nm
Connection head material
Aluminium
Stainless steel
1/2 NPT
32
35
3/4 NPT
36
40
M20 x 1.5, with counter nut
1)
23
25
M24 x 1.5, with counter nut
1)
27
30
1) only for versions with non-divisible neck tube
Tightening torques between connection head and neck tube
Tightening torques for connection to thermowell
Tightening torques for connection to neck tube
Thread
Tightening torques in Nm
R 1/2
1)
50 ... 60
1) only for versions with “nipple-union-nipple” neck tube
Thread
Tightening torques in Nm
1/2 NPT
35
3/4 NPT
40
G 1/2 B
35
G 3/4 B
40
M14 x 1.5
25 ... 30
M18 x 1.5
35
M20 x 1.5
35 ... 40
M27 x 2
40 ... 45
■
Only ever screw in, or unscrew, the instrument via the spanner-flats and to the prescribed
torque using an appropriate tool.
■
The correct torque depends on the dimensions of the connection thread and the gasket
used (form/material).
■
When screwing in or unscrewing the instrument, do not use the connection head as
contact surface.
■
When screwing in the instrument, please observe that the threads are not skewed.
4.6 Scope of delivery
Cross-check scope of delivery with delivery note.
4.4 Housing- and connection heads
The dimensions of the housing and connection heads are given in the respective data sheet.
4.5 Tightening torques
4. Design and function