7. Maintenance and cleaning

7.1 Maintenance

The thermometers described here are maintenance-free.

Repairs must only be carried out by the manufacturer.

7.2 Cleaning

CAUTION!

Prior to cleaning the electrical connections, disconnect them properly.

Clean the instrument with a moist cloth.

Electrical connections must not come into contact with moisture.

Wash or clean the dismounted instrument before returning it, in order to

protect persons and the environment from exposure to residual media.

Residual media in the dismounted instrument can result in a risk to persons,

the environment and equipment. Take sufficient precautionary measures.

For information on returning the instrument see chapter 9.2 “Return”.

7.3 Calibration, Recalibration

It is recommended that the measuring insert is recalibrated at regular intervals (resistance

thermometers: approx. 24 months, thermocouples: approx. 12 months). This period can be

reduced, depending on the particular application. The calibration can be carried out by the

manufacturer, as well as on site by qualified technical staff with calibration instruments.