4 Design and Function

10

WIKA Operating Instruction, Model CPG500

14034181.

01

01

/201

3

G

B/

D

GB

Key

Description

Turns the device on / off.
Press for 2 seconds. Turns on the backlight (stays on for 20 seconds).

Selects display unit: MIN, MAX or FS Minimum value Pressure peak

Displays the upper limit of the scale (e.g. 700 bar)

Press for 2 seconds. Changes the unit.
Auto Power Off – on/off. Zero point calibration.

Erases MIN and MAX values from the memory.
Confirms the MENU functions.

4.4

Menu functions


The following settings can be made in the MENU function:

Auto Power Off – on or off

Unit selection (bar, PSI, kPa, MPa)

Press the MENU key for 2 seconds to activate the functions menu. Press the MENU key again to
select the next function.
Press the OK key to save the function setting.
The device then switches to the display mode.

4.5

Connection to the process


The CPG500 digital pressure gauge is available with male thread G1/4 or G1/2 for the
corresponding versions.



When fitting the CPG500 digital pressure gauge do not exceed the specified
torques.

Pressure connection Torque

G ¼

25 Nm

G ½

25 Nm




When fitting directly, ensure that the case of the CPG500 digital pressure gauge can
be rotated freely.