6 Commissioning, operation

WIKA Operating Instruction, Model CPG500 13

14034181.

01

01

/201

3

G

B/

D

GB

6.2

Operating the CPG500 digital pressure gauge

6.2.1 Turning on (ON)


















6.2.2 Turn off (OFF)

Press once (briefly)

6.2.3 Turn on backlight

Press

for 2 seconds. The backlight is shut off automatically after 20 seconds.


6.2.4 MIN/MAX indication


Use this key to toggle the required value.
The key function is sequential; the values are indicated in the display in sequence.

The MIN/MAX function is used to measure pressure peaks. The respective lowest (MIN) and

highest (MAX) measured values are stored in the MIN/MAX memory. Values in the MIN/MAX

memory are erased when the device is turned off. If different pressure tests are to be carried out in

succession, the MIN/MAX memory must be erased after each measurement.

MIN/MAX-value is indicated in the display.

A self-test procedure is carried out

The measuring range is indicated (FS)

Auto Power Off function is active.
Power off activates automatically after
5 minutes. This function can be
altered in MENU.

Display mode:
ACT value displayed