Hand test pump

GB

CPP 700-H / CPP 1000-H


WIKA Operating Instructions Hand test pump Version 1.3

6

3. Mounting instructions

.


The reference instrument is fitted to the upper side of the hand test pump CPP 700-H or

CPP 1000-H. The reference instrument is sealed by the integral O-ring sealing gasket. The

maximum torque is 15 Nm.


The unit under test is mounted to the end of the flexible test hose. Tighten to the connector to

prevent any leaks to a maximum torque of 15 Nm.


Never apply external pressure to the CPP 700-H and CPP 1000-H. Do not
connect to external pressure sources.




As an accessory, several stainless steel sets of adapters are available to cover different thread

dimensions of your test specimen.


Tighten the optional adapters to a maximum torque of 15 Nm.


Tip: It is possible to connect the test specimen directly to the pump. After dismounting
of the hose connector, there is a 1/4" BSP female thread at the side of the pump body.




Open the filling plug (3) of the liquid reservoir (6) and fill with a suitable fluid:

mineral oil based hydraulic fluid or clean water, free of calcium-carbonate / scale.

(Not suitable for hydraulic oils based on water and destilled water.)


Fill the liquid reservoir (6) not completely full